Patients doses Estimation in Common Diagnostic X-Ray Examinations and Local DRLs Determination for All Types of Patients
Hossein Rahanjam Mr1, Hossein Gharaati Dr * 2, Mohamadreza Kardan Dr3, Behzad Fasaee Mr4, Arash Akbarzadeh Mr5
1- MSc MSc Student of Radiobiology and Radiation Protection, Radiology Department, School of Paramedical Sciences, Tehran University of Medical Science, Tehran, Iran
2- PhD Master of Medical Physics and Radiotherapy Oncology Department, Medical Physics Department and Cancer Research Center, Tehran University of Medical Sciences, Tehran, Iran , hgharaati@yahoo.co.uk
3- phD Master of Nuclear Science and Technology Research Institute, Nuclear Safety and Radiation Protection Research Group, Atomic Energy Organization of Iran, Tehran, Iran
4- MSc MSc Degree of Medical Physics, Nuclear Safety and Radiation Protection Research Group, Atomic Energy Organization of Iran, Tehran, Iran
5- MSc . MSc student of Biostatistics, Epidemiology and Biostatistics department, school of Public Health, Tehran University of Medical Science, Tehran, Iran
Abstract:   (1275 Views)

Objective: Nowadays medical imaging is essential in any medical diagnosis. In 1990, the International commission on radiological protection recommended diagnostic reference level (DRL)  to optimize the X-ray examinations according to the principle of optimization assisting the patient radiation dose in diagnostic imaging centres not being ignored.

Material and Method: 12 common radiography examinations have been conducted for all type of patients in all of Ilam hospitals and imaging centres. The quality control check was performed on the equipment and it used the solid state dosimeters for estimating the Entrance Surface Air Kerma. Then Entrance Skin Dose of patient was estimated. Finally, its third quartile (75%) was presented as DRL.

Result: The range of Entrance Skin Dose (ESD) is from 18 mGy to 0.001 mGy in lumbosacral lateral (LAT) in fat adult and Chest of an infant, respectively. Also the range of DRL is from 4.62 mGy is as low as 0.07 mGy in lumbosacral of an adult and infant’s chest, respectively.

Conclusion: All DRLs obtained in this study are compared with other reference, countries and other cities in Iran. Almost DRLs in Ilam are less than them. One of the reasons may due to the execution of QA program. Establishing DRL in a province and encouraging radiographers to follow these values; carry out QA program regularly in all radiology departments will causes more effective radiation protection in population and patient dose.

Keywords: Entrance Skin Dose, Diagnostic Reference Level, Radiography Examination, Quality Control
Full-Text [PDF 321 kb]   (930 Downloads)    
Article Type: Original Contributions | Subject: Health Management
Received: 2016/03/2 | Accepted: 2016/08/22 | Published: 2016/10/15
Send email to the article author

Add your comments about this article
Your username or Email:

Write the security code in the box >



XML     Print


Download citation:
BibTeX | RIS | EndNote | Medlars | ProCite | Reference Manager | RefWorks
Send citation to:

Rahanjam H, Gharaati H, Kardan M, Fasaee B, Akbarzadeh A. Patients doses Estimation in Common Diagnostic X-Ray Examinations and Local DRLs Determination for All Types of Patients. CJHR. 2016; 2 (1) :44-53
URL: http://cjhr.gums.ac.ir/article-1-50-en.html
Volume 2, Issue 1 (9-2016) Back to browse issues page
Caspian Journal of Health Research
 
Creative Commons License
Persian site map - English site map - Created in 0.15 seconds with 31 queries by YEKTAWEB 3638